i-Series - Specs
LC-2050C (UV model)
Pump
| Degassing unit | 
                                   Five Lines: Mobile phase 4 + Rinse solution 1 [Volume 400 μL]  | 
                              
|---|---|
| Pumping method | 
                                   Parallel-type double plunger  | 
                              
| Pulsation | 
                                   ≤ 0.1 MPa [1.0 mL/min,10 MPa, Water]  | 
                              
| Flow rate setting range | 
                                   0.0001 to 10 mL/min  | 
                              
| Flow rate accuracy | 
                                    ≤ ±1% or ≤ ±2 μL/min, below whichever is greater  | 
                              
| Flow rate precision | 
                                   < 0.06 %RSD or < 0.02 minSD, below whichever is greater  | 
                              
| Configuration | 
                                   Four-solvent low-pressure gradient  | 
                              
| Gradient / range of set concentrations | 
                                   0 to 100%, in 0.1% steps  | 
                              
| Gradient / concentration accuracy | 
                                   ±0.5% [0.1 to 2 mL/min, 1 to 20 MPa, Specified condition]  | 
                              
| Gradient / concentration precision | 
                                   ±0.1% [1 mL/min,10 MPa, Specified condition]  | 
                              
| Maximum pressure | 
                                   50 MPa [0.0001 to 5 mL/min]  | 
                              
| System Delay Volume | 
                                   650 μL  | 
                              
Autosampler
| Sample injection method | 
                                   Total-volume sample injection  | 
                              
|---|---|
| Injection volume accuracy | 
                                   ±1% [50 μL, N = 10]  | 
                              
| Injection volume setting range | 
                                   0.1 to 100 μL  | 
                              
| Injection volume reproducibility | 
                                   RSD < 0.20% [5.0−2000 μL]  | 
                              
| Cross-contamination | 
                                   0.0025% [Caffeine, Specified condition]  | 
                              
| Injection cycle time | 
                                   Min. 14 sec [Specified condition]  | 
                              
| Samples for processing | 
                                   336 [1 mL], 216 [1.5 mL], 112 [4 mL], 4 [MTP/DWP]  | 
                              
| Sample cooler | 
                                   4 to 45°C  | 
                              
| Injection linearity | 
                                   > 0.9999  | 
                              
Column Oven
| Heating and cooling method | 
                                   Forced air circulation method  | 
                              
|---|---|
| Containable column size | 
                                   6 pieces at 10 cm max., 3 pieces at 10 to 30 cm  | 
                              
| Temperature control range | 
                                   Room temperature − 12 to 90°C, Setting range 4 to 90°C  | 
                              
| Temperature control precision | 
                                   ±0.1°C  | 
                              
| Temperature stability | 
                                   ±0.8°C (Specified condition)  | 
                              
| Flow rate switching valve | 
                                   Max. 1 pc  | 
                              
UV Detector
| Wavelength range | 
                                   190 to 700 nm  | 
                              
|---|---|
| Spectral bandwidth | 
                                   8 nm  | 
                              
| Wavelength accuracy | 
                                   ≤ ±1 nm  | 
                              
| Wavelength Reproducibility | 
                                   ≤ ±0.1 nm  | 
                              
| Noise | 
                                   ≤ ±2.5 × 10−6 AU, (250 nm, Specified condition)  | 
                              
| Drift | 
                                   ≤ 100 × 10−6 AU/h [250 nm, Specified condition]  | 
                              
| Simultaneous monitoring of 2 wavelengths | 
                                   Enable [Any 2 wavelengths of 190 to 370 nm or 371 to 700 nm]  | 
                              
| Linearity | 
                                   Up to 2.5 AU (5%)  | 
                              
| Sampling rate | 
                                   Up to 100 Hz  | 
                              
| Light source | 
                                   Deuterium (D2) lamp  | 
                              
| Flow Cell | 
                                   12 μL (10 mm, TC), 12 MPa  | 
                              
| Option cell | 
                                   High-Speed: 8 μL [10 mm, TC]  | 
                              
Miscellaneous
| Dimensions | 
                                   W410×H605×D500 mm [Not including reservoir tray height]  | 
                              
|---|---|
| Weight | 
                                   63 kg  | 
                              
| Available pH range | 
                                   1 to 13  | 
                              
| Materials for parts in contact with liquids | 
                                   Stainless steel (SUS316L, SUS316), FEP, PEEK, PTFE, perfluoroelastomer, ruby, sapphire, Hastelloy® C, GFP, ceramic, PFA, quartz, PPS  | 
                              
Downloads
Download the latest brochure.